PRODUCTS
VIBRATION
 Features

TV RANGE
Single-stage thermostatic chambers with vibration device – temperature range –40°C + 180°C

  • Light alloy frame, painted in RAL 9010;
  • Completely removable aluminium finishing panels, painted in RAL 9010;
  • Fire-proof (MUPAM) glass wool thermal insulation
  • Internal cell made of 18/8 AISI 304 stainless steel with 15/10 thickness;
  • Forced air circulation obtained by 1 fan with a +70°C resistant motor;
  • PID electronic regulator with temperature digital display and with RS 422 serial interface, which can be monitored from a compatible PC;
  • High and low temperature fail-safe electronic thermostat;
  • High temperature fail-safe mechanical thermostat, connected to an automatic thermostat with spring release coil
  • Door safety switch
  • Emergency button;
  • Electronic system with moulded modular components to make possible maintenance operations easier and quicker;
  • All instruments are placed in the electric control board on the front of the chamber.

PERFORMANCE

TEMPERATURE

  • Range: - 40°C to + 180 °C;
  • Precision: ± 1 °C after stabilisation;
  • Heat-up and cool-down rate: 5 °C/min (linear degrees) in temperature range of – 30° to 180°C

TV RANGE
Two-stage thermostatic chambers – temperature range -80°C + 180°C

PERFORMANCE

TEMPERATURE

  • Range: - 80°C to + 180 °C;
  • Precision: ± 1 °C after stabilisation;
  • Heat-up and cool-down rate: 5 °C/min (linear degrees) up to – 54° to 180°C
  • OPTION: linear heat-up and cool-down rate 7, 10°C/min up to -54°C

THV RANGE
Single and two-stage climatic chambers with vibration device – temperature range –40°C / - 80°C + 180°C

  • PID electronic regulator with relative humidity digital display and with RS 422 serial interface – which can be monitored from a compatible PC;
  • Electronic capacitive probe to determine relative humidity as a percentage;

HUMIDIFICATION SYSTEM

  • The humidification system will be produced by a steam generator, which is directly piloted by the control panel;
  • Polyethylene tank for supplying the steam generator with distilled water, so as to avoid limescale deposits on the components tested.

DEHUMIDIFICATION SYSTEM

  • Dehumidification will take place through the condensation of humidity on a specific large refrigerated battery, in order to assure dehumidification gradient volumes.
    The chamber will be built so that the condensate is automatically drained off the chamber.

PERFORMANCE

TEMPERATURE

  • Range: - 80°C to + 180 °C;
  • Precision: ± 1 °C after stabilisation;
  • Heat-up and cool-down rate: 5 °C/min (linear degrees) up to – 54° to 180°C.

RELATIVE HUMIDITY

  • Range: from 10% to 98% of climatic range from 10°C to 95°C with dew-point limitation
  • Precision: ± 0,5% ± 2%;