TA RANGE
Single-stage thermobaric chambers – temperature range –40°C + 150°C
- Light alloy frame, painted in RAL 9010;
- Completely removable aluminium finishing panels, painted in RAL 9010;
- Fire-proof (MUPAM) glass wool thermal insulation;
- Internal cell made of 18/8 AISI 304 stainless steel with 30/10 thickness;
- Chamber reinforcements with rectangular tubes made of 30/10 thick 18/8 AISI 304 stainless steel;
- Forced air circulation obtained by 1 or more fans with blades with external extended axis motor equipped with automatically lubricated sealing ring;
- PID electronic regulator with temperature digital display and with RS 422 serial interface, which can be monitored from a compatible PC;
- PID electronic regulator with vacuum digital display and with RS 422 serial interface, which can be monitored from a compatible PC;
- High and low temperature fail-safe electronic thermostat;
- High temperature fail-safe mechanical thermostat, connected to an automatic thermostat with spring release coil;
- Door safety switch;
- Emergency button;
- Electric system with moulded modular components to make possible maintenance operations easier and quicker;
- SAll instruments are placed in a drawer on the front of the chamber.
PERFORMANCE
TEMPERATURE
- Range: - 40°C to + 150 °C;
- Precision: + 0,5 °C after stabilisation;
- Heat-up and cool-down rate: average of 2 °C/min in temperature range of – 40° to 150°C
- OPTION: linear heat-up and cool-down rate 5, 7, 10°C/min up to -30°C
VACUUM
- Range: from 1000 mBar to 0 mBar in 15 minutes;
- Precision: + 3 mBar.
TA RANGE
Two-stage thermobaric chambers – temperature range -70°C + 150°C
PERFORMANCE
TEMPERATURE
- Range: - 70°C to + 150 °C;
- Precision: + 0,5 °C after stabilisation;
- Heat-up and cool-down rate: average of 2 °C/min in temperature range of – 70°C to 150°C
- OPTION: linear heat-up and cool-down rate 5, 7, 10°C/min up to -54°C.
VACUUM
- Range: from 1000 mBar to 0 mBar in 15 minutes;
- Precision: + 3 mBar.
THA RANGE
Single and two-stage humidity control thermobaric chambers – temperature range –40°C / - 70°C + 150°C
- PID electronic regulator with relative humidity digital display and with RS 422 serial interface, which can be monitored from a compatible PC;
- Electronic capacitive probe to determine relative humidity as a percentage;
HUMIDIFICATION SYSTEM
- The humidification system is operated by a steam generator, which is directly monitored by the control device;
- Polyethylene tank for supplying the steam generator with distilled water, so as to avoid limescale deposits on the components tested;
DEHUMIDIFICATION SYSTEM
- Dehumidification will take place through the condensation of humidity on a specific large refrigerated battery, in order to assure dehumidification gradient volumes.
The chamber is built so that the condensate is automatically drained off the chamber.
PERFORMANCE
RELATIVE HUMIDITY
- Range: from 10% to 98% of climatic range from 10°C to 95°C with dew-point limitation
- Precision: + 0,5% + 2%;
Serie THA
with LCD PC and touch screen |
|
Serie TCA
Cooling System realized for collaudi di satelliti
|
|
 |
Serie THA
-70°C to + 150°C |
 |
|
|